Spectranetics Corporation LLD EZ Lead Locking Device; Stylet, Catheter Back to Search Results
Model Number 518-062
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 08/17/2026
Type of Reportable Event Serious Injury
Event or Problem Description
A lead extraction procedure commenced to remove a right atrial (ra), right ventricular (rv), and left ventricular (lv) lead due to bacteremia.Spectranetics lld ez lead locking devices (lld ezs) were inserted into each lead to provide traction.Beginning with a spectranetics 14f glidelight laser sheath, along with a spectranetics visisheath dilator sheath, advancement was made through the subclavian down past the ra.With traction from the lld ez, the lv lead was extracted; however, a pericardial effusion was detected.Rescue efforts began, including bypass, pericardial window, and sternotomy.A perforation to the coronary sinus (cs) was discovered and repaired (mdr #3007284006-2026-00431).The decision was made to abandon the extraction; therefore, the lld ezs, along with the ra and rv lead, were cut and capped and remained in the patient (mdr #3007284006-2026-00432).There was no attempt made to unlock the lld ezs.The patient survived the procedure.This report captures the lld ez within the ra lead, which was cut/capped and remained in the patient.There was no alleged malfunction of any spectranetics devices in use during the procedure.
 
Additional Manufacturer Narrative
A3b) patient gender - not available from facility.H3) a portion of the device was discarded, and a portion remained within the lead within the patient.H6) lld cut/cap is a known risk of complication with use of the lld.Although lld cut/cap is a known risk of complication with use of the lld, the physician did not attempt to unlock the lld ez from the lead prior to cutting and capping.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LLD EZ Lead Locking Device
Common Device Name
Stylet, Catheter
Manufacturer (Section D)
Spectranetics Corporation
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
Spectranetics Corporation
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
nicole worford
9965 federal drive
colorado springs, CO 80921
+1(719)447-2465
MDR Report Key50498457
Report Number3007284006-2026-00433
Device Sequence Number16005322
Product Code DRB
UDI-Device Identifier00813132023072
UDI-Public(01)00813132023072(17)280309(10)0303726156
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K142116
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number518-062
Device Catalogue Number518-062
Device Lot Number0303726156
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2026
Initial Report FDA Received Date08/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Medtronic 4598 LV lead.; Medtronic 5076 RA pacing lead.; Medtronic 6935M RV ICD lead.; Spectranetics 14F GlideLight Laser Sheath.; Spectranetics CVX-300 Excimer Laser System.; Spectranetics LLD EZ Lead Locking Devices.; Spectranetics VisiSheath Dilator Sheath.
Patient Age46 YR
Patient SexMale
Patient Weight130 LB
Patient RaceWhite
-
-