Updated fields: d4, d9, g3, g6, h2, h3, h6, h11
the hakim programmable valve (id 823100) was not returned for evaluation (not available for return as per customer) and lot number information has not been provided; therefore, an evaluation of the device could not be performed, and device history record (dhr) could not be reviewed.The root cause(s) of the reported issue could not be determined, however, the possible root cause for the issue reported by the customer could be mechanical stress applied to the white marker during implantation, handling, pressure setting adjustments, or in vivo use, leading to detachment of the marker.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|