• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DePuy Ireland - 3015516266 KINCISETM; Power Tool Adaptor

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DePuy Ireland - 3015516266 KINCISETM; Power Tool Adaptor Back to Search Results
Catalog Number 200015005
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the adaptor does not lock into kincise impactor.The issue was observed during surgery with delay there were no reports of injuries, medical intervention or prolonged hospitalization.It was reported that there was no further information regarding the issue.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary the device associated with this report was returned to depuy synthes for evaluation.The depuy synthes team forwarded the device to the service and repair site which conducted a visual and functional inspection of the returned device.The kincise 1 emphsys strght impct was visually and functionally assessed and determined to have failed the tool coupling assessment due to the locking collar fitting (stem deformed) damage, which prevented it from being fully inserted into the impactor, consistent with the adapter not being properly secured ¿ user error.No further action is required at this time since the reported issue is consistent with improper handling - user error.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the kincise 1 emphsys strght impct would have contributed to the complained device issue.¿ based on the investigation findings, it has been determined that no corrective and/or preventative action is required.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KINCISETM
Common Device Name
Power Tool Adaptor
Manufacturer (Section D)
DePuy Ireland - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
+1(908)386-3534
MDR Report Key50500057
Report Number1818910-2026-14766
Device Sequence Number13119676
Product Code HWE
UDI-Device Identifier10603295527848
UDI-Public(01)10603295527848(10)GM6469010
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
Exempt
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number200015005
Device Lot NumberGM6469010
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer05/18/2026
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 08/25/2026
Initial Report FDA Received Date08/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/05/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-