| Model Number |
6947M55 |
| Medical Device Problem Codes |
Fracture (1260); Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Continuation of d10: dtpa2d4 crt-d, implanted: (b)(6) 2026.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead exhibited an apparent fracture.It was also reported that the cardiac resynchro nization therapy defibrillator (crt-d) setscrew was noted to be sideways in the port, observed to be misaligned.In addition, there was high rv pacing impedance that was noted.However, further investigation noted that the rv lead performance was okay as it turned out that the crt-d set screw was misaligned.The crt-d was explanted and replaced, and the rv lead remains implanted active.No patient complications have been reported as a result of this event.
|
| |
|