MAUDE Adverse Event Report: Medtronic Europe Sarl Cobalt¿ XT DR MRI SureScan¿; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medtronic Europe Sarl Cobalt¿ XT DR MRI SureScan¿; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number DDPA2D1
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Sepsis (2067); Medical device site infection (4845)
Date of Event 07/20/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the right ventricular (rv) lead exhibited tip to coil out of range impedance that was noted to be chronically near the lower limit, suspected to be slightly below the alert threshold.It was further reported, via follow up, no intervention was performed.It was also noted the patient had a pocket infection and sepsis.The implantable cardioverter defibrillator (icd) system remains in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Cobalt¿ XT DR MRI SureScan¿
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
Medtronic Europe Sarl
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
Manufacturer (Section G)
Medtronic Europe Sarl
route du molliau 31
case postale
tolochenaz vaud 1131
SZ   1131
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
+1(763)526-9277
MDR Report Key50501081
Report Number9614453-2026-03469
Device Sequence Number10244973
Product Code LWS
UDI-Device Identifier00763000178338
UDI-Public(01)00763000178338(17)270414(20)02
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P980016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberDDPA2D1
Device Catalogue NumberDDPA2D1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2026
Initial Report FDA Received Date08/28/2026
Date Device Manufactured10/27/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5076-52 LEAD, 694965 LEAD
Outcome Attributed to Adverse Event Life Threatening;
Patient Age86 YR
Patient SexFemale
-
-