| Model Number |
SC-8216-50 |
| Medical Device Problem Codes |
Unexpected Therapeutic Results (1631); Migration (4003)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
01/16/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient experienced inadequate pain relief.The patient felt a pop and per physician the paddle spinal cord stimulation (scs) lead had migrated as confirmed via imaging.The patient underwent spinal cord stimulator (scs) explant procedure and was doing well postoperatively.All explanted device components were not released by the facility and will not be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional suspect medical device component involved in the event:
model number/catalog number: sc-1216.Serial number: (b)(6).Batch/lot number: 812015.Model/catalog description: wavewriter alpha 16 ipg kit.Unique identifier (udi): (b)(4).Investigation results:
the devices were not returned for analysis; therefore, a technical analysis could not be performed.Device history record:
it was confirmed these devices met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review:
review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Investigation conclusion:
based on a thorough review of the reported complaint, an investigation conclusion code of cause not established was assigned to the investigation for all components.
|
| |
|