| Model Number |
5392 |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Codes |
Arrhythmia (1721); Cardiac Arrest (1762); Ventricular Fibrillation (2130); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the external pulse generator (epg) incorrectly sensed the t wave and delivered pacing output.It was noted that the sense amplitude function was not working correctly.No patient complications have been reported as a result of this event.
|
| |
|
Event or Problem Description
|
|
It was further reported that the patient had epicardial wires connected to an external pulse generator (epg) and was maintained on a backup rate.The patient subsequently experienced an r on t event, followed by ventricular fibrillation and cardiac arrest.It was noted that the backup pacing settings prior to the arrest were noted as a rate of 50 beats per minute, sensitivity of 2 millivolts, output of 10 volts, and single chamber fixed rate (vvi) mode.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: analysis was unable to confirm the customer's comment that the external pulse generator (epg) incorrectly sensed the t wave and delivered pacing output.It was noted that the sense amplitude function was not working correctly.The device passed all incoming tests with no anomalies observed.Additionally, it was noted during the visual inspection that the hanger was cracked and all six screws were contaminated.All found defective parts were replaced, and all other identified issues were resolved.The epg then passed all final functional tests.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|