Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary.The product was returned to depuy synthes for evaluation.¿visual inspection of the returned device revealed damage to the aiming arm - center part and the latch, both of which were found broken.Additionally, the pin was separated from the device as a result of the damage observed in this area of the product.All device components were returned and available for evaluation.¿
the rfn-advanced retrograde femoral nailing system, was reviewed.Following relevant statements were found at page 24.Attach the aiming arm to the insertion handle, by sliding it into the hook at the distal part of the insertion
handle and then rotating the latch towards the insertion handle for both parts to connect.Precaution:
do not exert forces on the aiming arm, protection sleeve, drill sleeves and drill bits.These forces may prevent.Accurate targeting through proximal locking holes and damage the drill bits.A dimensional inspection was unable to be performed due to post manufacturing damage.¿a functional evaluation was unable to be performed due to post manufacturing damage.¿the overall complaint was confirmed as the observed condition of the aim-arm radioluc would have contributed to the complained device issue.¿
based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.¿
dimensional inspection: n/a.Device history.Per franchise complaint product investigation procedure complaints for which a non-manufacturing related probable cause has been identified no manufacturing record evaluation is required.
|