| Catalog Number |
4723 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/06/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that after the needle was first inserted into the meniscus, the first peek anchor was released.When the needle was withdrawn, the anchor followed the needle and the system was no longer usable.The anchor popped out of the meniscus without anchoring properly.The next product worked correctly.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation was previously closed based on product not received; however, the product has now been physically received at stryker endoscopy, usa and the investigation has been re-opened.Investigation as follows is now based on product received.Alleged failure: after the needle was first inserted into the meniscus, as recommended by the manufacturer, the first peek anchor was released.When the needle was withdrawn, the anchor followed the needle and the system was no longer usable.The anchor popped out of the meniscus without anchoring properly.The next product worked correctly.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause could be use of excessive force.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|