| Model Number |
SC-1110-02 |
| Medical Device Problem Code |
Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
Inappropriate Device Stimulation of Tissue (4928)
|
| Date of Event |
04/21/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the remote control kept on disconnecting from the device after a few minutes and patient does not like to feel the stimulation.The patient underwent an implantable pulse generator (ipg) replacement procedure and was doing well post operatively.No other information could be obtained.
|
| |
|
Additional Manufacturer Narrative
|
|
Block b3: approximated based on the date the manufacturer became aware of the event.Block d2b: pro code (product code): qrb.Investigation results:
the device was not returned for analysis; therefore, a technical analysis could not be performed.Device history record:
it was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review:
review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Media inspection:
photo provided by the customer did not provide relevant information to confirm the reported event.Investigation conclusion:
based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|
| |
|