• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO Puerto Rico Mfg. Inc. TECNIS IOL; Intraocular Lens

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO Puerto Rico Mfg. Inc. TECNIS IOL; Intraocular Lens Back to Search Results
Model Number DXW300
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Discomfort (2330)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a single-piece preloaded toric intraocular lens (iol) with a platinum cartridge was removed from the patient's left eye.The reason for the iol exchange was patient dissatisfaction with the quality of the implanted lens.The original lens was replaced with a zcu375, single-piece non-preloaded toric intraocular with a +20.0d lens on a secondary surgery.The device was not available for return.No further information was provided.
 
Additional Manufacturer Narrative
Section a4: weight: unknown/not provided section a5: ethnicity: unknown/not provided section a6: race: unknown/not provided section b3: date of event: unknown; requested but not provided.The best estimate date is between (b)(6)2026 "[implant date]" and (b)(6) 2022 "[explant date]".Device evaluation: product evaluation was not performed because the product has not been received.The complaint issue reported could not be confirmed.Manufacturing record review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search in the complaint system revealed that no other complaints were received for this production order number.Conclusion: based on the investigation, no malfunction or product deficiency was identified.Attempts were made to obtain the missing information; however, no additional information has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
Intraocular Lens
Manufacturer (Section D)
AMO Puerto Rico Mfg. Inc.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
+1(714)247-8552
MDR Report Key50506306
Report Number3012236936-2026-000252
Device Sequence Number10211657
Product Code HQL
UDI-Device Identifier05050474765306
UDI-Public(01)05050474765306(17)260819
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/19/2026
Device Model NumberDXW300
Device Catalogue NumberDXW300U185
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/03/2026
Initial Report FDA Received Date08/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/19/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-