| Catalog Number |
856207004 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
07/30/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during an initial fixation procedure, a screw guide became stuck to the plate and the screwdriver tip fractured during attempted removal.To confirm the screw insertion direction, the guide had been reattached to the plate.When removal was attempted, the guide could not be released using a screwdriver.Increased force was applied, causing the screwdriver tip to break, with a fragment remaining inside the guide.The guide could not be removed, and the procedure was completed with the guide still attached to the plate.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).D10:
item # 214215070, t-15 tapered driver, lot # rl303885.Item # unknown, unknown fast guide, lot # unknown.G2: foreign, event occurred in japan.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
| |
|