• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Zimmer Biomet, Inc. T-15 TAPERED DRIVER; Driver, Prosthesis

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Zimmer Biomet, Inc. T-15 TAPERED DRIVER; Driver, Prosthesis Back to Search Results
Catalog Number 214215070
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 07/30/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during an initial fixation procedure, a screw guide became stuck to the plate and the screwdriver tip fractured during attempted removal.To confirm the screw insertion direction, the guide had been reattached to the plate.When removal was attempted, the guide could not be released using a screwdriver.Increased force was applied, causing the screwdriver tip to break, with a fragment remaining inside the guide.The guide could not be removed and the procedure was completed with the guide still attached to the plate.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4) d10: item # 856207004, fib anatomic lock rt 4h ste, lot # 67503195 item # unknown, unknown fastguide, lot # unknown g2: foreign - event occurred in japan.The customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T-15 TAPERED DRIVER
Common Device Name
Driver, Prosthesis
Manufacturer (Section D)
Zimmer Biomet, Inc.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
Zimmer Biomet, Inc.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
+1(574)526-0384
MDR Report Key50506650
Report Number0001825034-2026-02612
Device Sequence Number15815030
Product Code HWR
UDI-Device Identifier00887868024931
UDI-Public(01)00887868024931(11)220405(10)RL303885
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number214215070
Device Lot NumberRL303885
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/03/2026
Initial Report FDA Received Date08/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/05/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
See H11 narrative
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-