| Catalog Number |
214215070 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
07/30/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during an initial fixation procedure, a screw guide became stuck to the plate and the screwdriver tip fractured during attempted removal.To confirm the screw insertion direction, the guide had been reattached to the plate.When removal was attempted, the guide could not be released using a screwdriver.Increased force was applied, causing the screwdriver tip to break, with a fragment remaining inside the guide.The guide could not be removed and the procedure was completed with the guide still attached to the plate.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4)
d10:
item # 856207004, fib anatomic lock rt 4h ste, lot # 67503195
item # unknown, unknown fastguide, lot # unknown
g2: foreign - event occurred in japan.The customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|