| Model Number |
iS3-US |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Hypersensitivity/Allergic reaction (1907); Inflammation (1932)
|
| Date of Event |
08/12/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that following a right eye (od) trabecular microbypass stent system procedure in conjunction with an intracameral implant procedure, the patient experienced persistent postoperative inflammation, and a possible metal allergy.Per report, the devices were explanted approximately two and a half (2.5) months postoperatively with no complications.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information:
the device is expected to be returned for evaluation but has not been received at glaukos evaluation center.Therefore, product testing on the actual device could not be performed.The device history record review of the manufacturing lot was performed and there were no non-conformities found to be related to the reported event.A review of the device labeling and associated risk documents was completed.Inflammation and allergic reaction are identified in the labeling as known inherent risks of trabecular micro-bypass stent procedure.The ifu adequately provides instructions for stent implantation, precautions, and warnings for the proper use and handling of the device.An adverse event appears to have occurred but does not appear to have been a problem with the device or the way it was used.Inflammation and allergic reaction are established risks associated with use of the device, which is clearly specified in the product's labeling.Based on the information received, the root cause of the reported event could not be conclusively identified.Mfr# reference: (b)(4).
|
| |
|
Search Alerts/Recalls
|