| Model Number |
IO-500 |
| Medical Device Problem Codes |
Overheating of Device (1437); No Flow (2991); Unexpected Shutdown (4019)
|
| Health Effect - Clinical Codes |
Tachycardia (2095); Low Oxygen Saturation (2477)
|
| Date of Event |
08/03/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient's unit was overheating and stopped producing oxygen.As a result, the patient was not receiving enough oxygen and it caused shortness of breath and their heart rate to increase.The patient was able to make it back to their house where they have a backup unit troubleshooting did not resolve the issue.The patient was sent a replacement unit and it is working for them.
|
| |
|
Additional Manufacturer Narrative
|
|
B3 date of event is estimated.The unit was returned for evaluation on (b)(6) 2026.The unit was returned with reported system hot, which was not confirmed upon inspection, but muffler was blocked and filled with dust; it causes low flow.Internal 02 reading measured 92%, while the external 02 reading measured 87%.And psa-10 that indicates column is contaminated with moisture; it contributed to the low oxygen level.Low flow product manifold duo to debris inside.Leak from cannula barb possible dropped unit.Housing & uip was dirty & scratched due to probable rough cleaning.
|
| |
|