• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. HD AUTOCLAVABLE CAMERA HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. HD AUTOCLAVABLE CAMERA HEAD Back to Search Results
Model Number OTV-S7PROH-HD-10E
Medical Device Problem Codes No Display/Image (1183); Poor Quality Image (1408); Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/17/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d8, d9, h2, h6, h11.The device was returned to olympus for inspection, and the reported failure was confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: it is dim due to faulty charged coupled device.The most probable cause of the malfunctions is traced to component failure.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
No additional information received from the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HD AUTOCLAVABLE CAMERA HEAD
Common Device Name
CAMERA HEAD
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50512548
Report Number3002808148-2026-30370
Device Sequence Number18817229
Product Code FET
UDI-Device Identifier04953170252914
UDI-Public04953170252914
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K955404
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 08/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOTV-S7PROH-HD-10E
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/17/2026
Initial Report FDA Received Date08/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/13/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-