Updated field: b4,g3, g6, h2, h6(type of investigation, investigation findings, investigation conclusions), h11.A getinge field service engineer (fse) evaluated the unit and was able to confirm the reported issue.The fse replaced the video receiver to lcd cable.Product passed all functional and safety tests per manufacturer specifications.The equipment has been delivered to the customer and is ready for clinical use.Based on the available information the effect of display failure on the cs100 is confirmed.The video receiver to lcd cable was replaced to resolve the issue.The defective part was not returned for failure analysis because the spare parts need to be archived.Therefore, a possible cause could not be determined.Fault logs were not provided.
|