| Model Number |
FITBONE |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Received information of an issue during implantation of a fitbone nail.As per reporter, during distal screw locking, the maximum available orthofix 4.0 mm locking screw was insufficient to achieve both secure fixation and maintain transport mechanism functionality.The surgeon implanted a third-party 4.0mm ø steinman-pin (steel).Allegedly, this led to a surgery delay of 60-90 minutes.The patient is stable and in good health.The receiver and control unit have been tested positive after implantation.No allegation of product malfunction was reported by the user.
|
| |
|
Additional Manufacturer Narrative
|
|
No allegation of product malfunction was reported by user.Precautions from relevant instructions for use:
fitbone devices should be used together with their corresponding fitbone implants, components, accessories and instrumentation following the operative technique recommended by the manufacturer.Orthofix does not guarantee the safety and effectiveness of the fitbone when used in conjunction with other orthofix devices if not specifically indicated in the operative technique.
|
| |
|