| Model Number |
1201 |
| Medical Device Problem Codes |
High impedance (1291); Communication or Transmission Problem (2896); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/03/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient with a snm tined lead experienced intermittent device shutoff and observed a red indicator light on the remote control.Troubleshooting was performed, during which open impedances were identified.Additional information received indicated that the lead was bent.The patient subsequently underwent a replacement procedure.No further patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Block d2b:\
additional pro code: qon.\
\
block g4:\
pma number: p190006.\
\
concomitant product:\
model number/catalog number: 4101.\
serial number: (b)(6).\
batch/lot number: ax1t052526.\
model/catalog description: neurostimulator.\
unique identifier (udi) #: (b)(4).
|
| |
|