• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical (CRM-Sylmar) Quadra Allure MP CRT-P; Pulse Generator, Pacemaker, Implantable, With Cardiac Resynchronization (Crt-P)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical (CRM-Sylmar) Quadra Allure MP CRT-P; Pulse Generator, Pacemaker, Implantable, With Cardiac Resynchronization (Crt-P) Back to Search Results
Model Number PM3562
Medical Device Problem Code Unexpected Mode Switch (4076)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the pacemaker was found in back-up mode prior to the implant procedure.The device was not implanted.There was no patient involved.
 
Additional Manufacturer Narrative
The reported event of device in unexpected mode switch was confirmed.The device was received in backup mode with normal telemetry communication.Analysis of the device image indicated the device went to bvvi due to hardware reset error.The device was restored by reloading product code followed by analysis indicating normal interrogation results.Backup vvi mode was reproduced at lower temperature, and this is consistent with an integrated circuit anomaly.Assessment of total projected longevity was performed, and the device was found to have appropriate remaining longevity.A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Quadra Allure MP CRT-P
Common Device Name
Pulse Generator, Pacemaker, Implantable, With Cardiac Resynchronization (Crt-P)
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50515946
Report Number2017865-2026-18247
Device Sequence Number16122607
Product Code NKE
UDI-Device Identifier05414734510097
UDI-Public(01)05414734510097(10)A000184222(17)270731
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P030035
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/31/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPM3562
Device Lot NumberA000184222
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/13/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/10/2026
Initial Report FDA Received Date08/31/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/24/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-