| Model Number |
Mi1250 SYNCHRONY 2 PIN |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Wound Infection (2447)
|
| Date of Event |
02/18/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Symptoms leading to explantation were wound infection and implant extrusion.Reportedly, the user will decide in the future if he wants an implant.The electrode array was left in the cochlea.
|
| |
|
Event or Problem Description
|
|
Symptoms leading to explantation were wound infection and implant extrusion starting 3 months post op.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: device investigation did not reveal any device defect or damage, which has been present whilst implanted.According to the information received from the field the device was explanted due to a an infection and extrusion of the device.Review of the device_s sterilization records show that the device has been subject to a valid sterilization process, thus no available information points to the implant being the source of the infection.Additionally, there was a slight dislocation of the implant out of the implant bed, which might have contributed to the infection and extrusion.This is a final report.
|
| |
|