• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medtronic, Inc. Dual Chamber Temporary Pacemaker; PULSE-GENERATOR, PACEMAKER, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medtronic, Inc. Dual Chamber Temporary Pacemaker; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 5392
Medical Device Problem Codes Calibration Problem (2890); Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/31/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that one of the knobs of the external pulse generator (epg) was physically damaged.The epg has been returned for evaluation and subsequently tested out of specification during manufacturer's analysis.There was no patient involvement.
 
Additional Manufacturer Narrative
Product analysis: analysis was unable to confirm the customer's comment that one of the knobs of the external pulse generator (epg) was physically damaged.However, it was noted that the epg failed incoming tests for the rate control knob.It was also noted that the backlight on-off difference test failed.Additionally, one control knob was found detached from the epg but was not damaged.The upper case was observed to be broken around the rate encoder, and the hanger could not be folded into the case.The epg was returned unrepaired due to the expiration of its service life.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Dual Chamber Temporary Pacemaker
Common Device Name
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
Medtronic, Inc.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
Medtronic, Inc.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
+1(763)526-9277
MDR Report Key50517822
Report Number2182208-2026-50456
Device Sequence Number21502364
Product Code DTE
UDI-Device Identifier00643169230859
UDI-Public(01)00643169230859(11)140514
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K150246
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/31/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/07/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2026
Initial Report FDA Received Date08/31/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/14/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-