| Model Number |
4968-35 |
| Medical Device Problem Codes |
Failure to Capture (1081); Capturing Problem (2891); High Capture Threshold (3266)
|
| Health Effect - Clinical Code |
Heart Failure/Congestive Heart Failure (4446)
|
| Date of Event |
07/01/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the left ventricular (lv) epicardial lead began to exhibit rising/high thresholds. the device was reprogrammed at a higher output.The thresholds continued to rise and then the lv lead had no capture in any vector.The patient had worsening heart failure as a result.The lead was abandoned and replaced. no further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: date is approximate.Month and year are confirmed valid.Continuation d10: 407652 lead implanted (b)(6) 2020; 407645 implanted (b)(6) 2020 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|