| Model Number |
30001004101110012 |
| Medical Device Problem Code |
Failure to Deliver Shock/Stimulation (1133)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device failed to discharge.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation evaluated the device and the customer's report was not replicated or confirmed.The device was put through extensive testing including impedance calibration testing, defib/pacer stress testing, bench handling, defib cycling and all functional testing without duplicating the report.The device was found to be operating as intended.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
| |
|