Model Number 470500-13
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/06/2026
Type of Reportable Event Other
Event or Problem Description
Refer to h11 for follow-up information.
 
Additional Manufacturer Narrative
Device evaluation: the seal was analyzed and found to have a tear on the rubber duckbill measuring approximate 5.0mm in length.As result of the fully tearing, a piece measuring approximate 3.0mm x 2.0mm was detached and was returned with the sterile seal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
da Vinci
Common Device Name
SEAL
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
+1(408)523-2100
MDR Report Key50520070
Report Number2955842-2026-37437
Device Sequence Number9522429
Product Code GCJ
UDI-Device Identifier10886874116408
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K181395
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 08/31/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Operator of Device Health Professional
Device Model Number470500-13
Device Catalogue Number470500
Device Lot NumberK12251226 NA
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/27/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/06/2026
Initial Report FDA Received Date08/31/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/26/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-