(b)(4)
date sent: 9/23/2026
d4: batch #a97r9p.Additional information: d4 (lot number, expiration date), h4
correction: d4 (udi).Investigation summary:
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one ech60l device was returned with no apparent damage and with no reload present.The device was tested for functionality in the articulated position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples met the staple release criteria.The event described could not be confirmed as the device performed without any difficulties noted.The articulation mechanism was totally functional during functional testing.The device event log was reviewed the log indicates that no suspect power cycles were noted.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number a97r9p, and no non-conformances were identified.
|