Model Number 310MN33
Medical Device Problem Code Incomplete Coaptation (2507)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/14/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Additional information received reported that the 33mm mosaic neo bioprosthetic valve malfunctioned.It was reported that the mosaic neo bioprosthetic valve was fully sutured and tested prior to removal.No additional adverse patient effects were reported.
 
Event or Problem Description
It was reported that during the implant of this 33mm mosaic neo bioprosthetic valve, it was explanted and replaced with a non-medtronic product.The reason for the replacement was reported as poor coaptation of the valve leaflets.It was further reported that abnormal leaflet movement was noted when checking the valve after implantation using an asepto syringe and red rubber catheter. it was noted that the case required longer clamp and bypass time to complete.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
B5 and d9 updated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Mosaic NEO Bioprosthetic Heart Valve
Common Device Name
REPLACEMENT HEART-VALVE
Manufacturer (Section D)
Medtronic Heart Valves Division
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
Medtronic Heart Valves Division
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
+353(091)708096
MDR Report Key50522106
Report Number2025587-2026-00974
Device Sequence Number13439278
Product Code DYE
UDI-Device Identifier00763000990596
UDI-Public(01)00763000990596(17)310414(20)01
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P990064
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number310MN33
Device Catalogue Number310MN33
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/02/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2026
Supplement Date Received by Manufacturer09/19/2026
Initial Report FDA Received Date08/31/2026
Supplement Report FDA Received Date09/24/2026
Date Device Manufactured04/15/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age72 YR
Patient SexFemale
Patient Weight146 LB
-
-