| Model Number |
8637-20 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Scar Tissue (2060)
|
| Date of Event |
08/17/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was receiv ing dilaudid (hydromorphone) (dilaudid 50 mg/ml at 10.975 mg/day) via an implantable pump for non-malignant pain.It was reported that the healthcare provider (hcp) replaced the catheter and explanted the pump.When implanting the new pump, the pocket site had excessive bleeding and they did cautery to attempt to find source of bleeding.The hcp had a hard time finding where the bleeding was coming from due to scar tissue build up surrounding the old pump pocket.They were able to locate and purse string suture off vein but patient lost about 1l of blood.Hcp opted to cut and tie off new catheter he had just implanted and not implant new pump to let pocket heal.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|