It was reported that, during an unknown surgery, the first firing was completed without issue.However, during the second firing, the firing knob did not move.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
|
|
(b)(4).Date sent 8/31/2026.D4 batch # 850d65.Investigation summary:
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the ntlc75 device was received with no apparent damage and with a (b)(6)reload no loaded in the device.In addition, an opened package was received along with the device.The reload had the proximal 2 drivers up and the remaining drivers down with staples present and the swing tab in the locked position.In addition, the gripping surface from left side was noted to be damaged.The device was tested with the test reload and fired without any difficulties.The staple line was complete, and the staples meet the staple form release criteria.The event reported was confirmed and it is related to improper use of the device.It should be noted that the reload is designed to lockout, as a safety feature, if any staples have been fired from the cartridge reload.In addition, failure to complete the stroke may result in an incomplete staple line.This condition is consistent with an improper loading technique.When loading the cartridge in the device, make sure the cartridge is inside the channel track and the proper loading technique.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 850d65, and no non-conformances were identified.
|