| Model Number |
CF-HQ190L |
| Medical Device Problem Codes |
Erratic or Intermittent Display (1182); No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the colonovideoscope to the service center for evaluation related to a separate issue.During inspection, the service repair technician identified that the device had displayed no image and a black screen occurs intermittently.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: due to damage connector block and component failure.Three attempts were performed to obtain additional information, but no response was received from the customer.Date of event is unknown.Additional information will be provided if available.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|