| Model Number |
R1032804 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A trilogy 200 ventilator was returned to the third-party service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.There was no allegation of serious or permanent harm or injury.
during the evaluation of the device at a third-party service center, the device was visually inspected and found no evidence of foam degradation.However, an error code indicating a batt_low_state_of_health_int_li_ion failure occurred was observed in the device event log.No repairs were performed and the device will be scrapped.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported failure of internal battery is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.Device exceeds expected life.Dhr not required.
|
| |
|
Search Alerts/Recalls
|