The reported issue was moisture seeping into the glove, seen in the palms or around the finger tips.The returned samples were tested in accordance with fda guidance for leak detection, and all three of the sizes passed the requirements for leakage defects; inspection records confirmed the products all passed the requirements for leakage defects, and tensile and elongation results exceeded fda and astm 6319 minimums.The reported defect was not confirmed with received sample.At this time medline believed the reported moisture detected could be perspiration that can build inside the gloves during use, and corrective action type was none.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|