| Model Number |
470057-11 |
| Medical Device Problem Code |
Unintended System Motion (1430)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported event.The fse was unable to replicate the reported issue.Both endoscopes passed system test drive.Also, universal surgical manipulator (usm)s passed selected friction tests.The system was verified and ready to use.
|
| |
|
Event or Problem Description
|
|
It was reported that during da vinci-assisted surgical procedure, site contacted intuitive technical support engineer (tse) to report the 30-degree camera was orienting itself 90 degrees from horizontal and the master tool manipulator (mtm) were following.Customer tried a new camera and the issue remained.A hard power cycle was done, and site reseated the patient side cart (psc) fiber cable.The camera then oriented properly.Surgeon took control of the instruments with tse on the phone and reported the robot was working normally.Site was continuing with the case.The procedure was completed with no reported injury.
|
| |
|