It was reported that the procedure was to treat a lesion in the left anterior tibial artery.The 3x38mm esprit btk scaffold was advanced to the target lesion; however, the shaft kinked during a deployment attempt.Therefore, the device was removed from the patient.Another 3x38mm esprit device was used to continue the procedure.There was no adverse patient effect and no clinically significant delay reported in the procedure.No additional information was provided.
|
|
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the information received, the investigation determined that the reported issue appears to be related to operational context as it is likely handling and/or manipulation of the device during use resulted in the reported shaft kink.The kinked shaft likely caused a pinched/reduced inflation lumen causing the reported activation failure of the stent.Based on the results of the complaint investigation, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.C3: 1 per year was entered; however, the frequency is 1 dose per implant.
|