ABBOTT VASCULAR INC. ESPRIT BTK; Scaffold,infrapopliteal,absorbable Back to Search Results
Model Number 1203300-38
Medical Device Problem Codes Deformation Due to Compressive Stress (2889); Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/12/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was to treat a lesion in the left anterior tibial artery.The 3x38mm esprit btk scaffold was advanced to the target lesion; however, the shaft kinked during a deployment attempt.Therefore, the device was removed from the patient.Another 3x38mm esprit device was used to continue the procedure.There was no adverse patient effect and no clinically significant delay reported in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the information received, the investigation determined that the reported issue appears to be related to operational context as it is likely handling and/or manipulation of the device during use resulted in the reported shaft kink.The kinked shaft likely caused a pinched/reduced inflation lumen causing the reported activation failure of the stent.Based on the results of the complaint investigation, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.C3: 1 per year was entered; however, the frequency is 1 dose per implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESPRIT BTK
Common Device Name
Scaffold,infrapopliteal,absorbable
Manufacturer (Section D)
ABBOTT VASCULAR INC.
3200 lakeside drive
santa clara CA 95054 -280
Manufacturer (Section G)
ABBOTT VASCULAR, Reg # 2024168
26531 ynez road
temecula CA 92591 -462
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591--462
+1(951)914-5259
MDR Report Key50525616
Report Number2024168-2026-03391
Device Sequence Number15909744
Product Code NXW
UDI-Device Identifier08717648357015
UDI-Public(01)08717648357015(11)260113(17)280112(10)6011361
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P230036
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 08/31/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1203300-38
Device Catalogue Number1203300-38
Device Lot Number6011361
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/06/2026
Initial Report FDA Received Date08/31/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/13/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-