Model Number 0644-021-01
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/07/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during preparation prior to a procedure, it was noticed that the fiber was broken near the connector.The procedure was completed by using another fiber, patient outcome was not reported.
 
Event or Problem Description
It was reported that during preparation prior to a procedure, it was noticed that the fiber was broken near the connector.The procedure was completed by using another fiber, patient outcome was not reported.
 
Additional Manufacturer Narrative
The fiber was not returned to our post market quality assurance laboratory; therefore, neither a visual nor microscopic analysis was performed.The reported allegation of fiber break was not confirmed.A labeling review was performed.The labeling review found that the product ifu (instructions for use) states: inspect the fiber for kinks, punctures, fractures, or other damage.If the fiber appears damaged, do not use the fiber; return it to the supplier for replacement and hold the fiber tip to a non-reflective surface, and ensure that a circular red/green spot appears.If the spot is not circular, cleave the fiber.If the spot is weak or not visible, discard the fiber or return it to the supplier for replacement.A risk review was completed and confirmed that the event of fiber break is defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.A ship history review and a dhr (device history review) could not be performed due to the fiber upn and lot could not be identified.Based on the information available, a conclusion code of cause not established was assigned to this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SlimLine SIS EZ
Common Device Name
Powered Laser Surgical Instrument
Manufacturer (Section D)
Lumenis Ltd
hakidma 6
yokneam 20692 04
IS  2069204
Manufacturer (Section G)
LUMENIS LTD
hakidma st 6
yokneam, 20692 04
IS   2069204
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50526949
Report Number2124215-2026-46110
Device Sequence Number17655024
Product Code GEX
UDI-Device Identifier07290109145600
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K170121
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Type of Report Initial,Followup
Report Date (Section B) 09/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0644-021-01
Device Catalogue Number0644-021-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/07/2026
Supplement Date Received by Manufacturer09/11/2026
Initial Report FDA Received Date08/31/2026
Supplement Report FDA Received Date09/21/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-