• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Biosense Webster Inc THERMOCOOL SMARTTOUCH SF; Cardiac Ablation Percutaneous Catheter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Biosense Webster Inc THERMOCOOL SMARTTOUCH SF; Cardiac Ablation Percutaneous Catheter Back to Search Results
Catalog Number D134805
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Cardiac Perforation (2513)
Date of Event 08/05/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent an idiopathic ventricular tachycardia ablation procedure with a thermocool smarttouch sf catheter and the patient experienced cardiac perforation which required pericardiocentesis and surgical intervention.During the procedure, 17 ablations were performed outside of the pulmonary veins.While ablating in the right ventricular outflow tract, the patient¿s blood pressure dropped.A pericardial effusion was diagnosed via intracardiac echocardiography (ice).Pericardiocentesis was performed by the physician.The patient was stable and was transferred to the operating room for further surgical intervention.Patient has fully recovered.Physician¿s opinion on the cause of the event is that there was a perforation during ablation or catheter manipulation.
 
Additional Manufacturer Narrative
Device investigation details: an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.However, if the product is received at a later date, the investigation will be updated as applicable.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Note: for field d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's ref.# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL SMARTTOUCH SF
Common Device Name
Cardiac Ablation Percutaneous Catheter
Manufacturer (Section D)
Biosense Webster Inc
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
Biosense Webster Inc (Juarez)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
katie seppa
31 technology dr
irvine, CA 92618
+1(513)337-8052
MDR Report Key50528618
Report Number3013300026-2026-02851
Device Sequence Number18619528
Product Code LPB
UDI-Device Identifier10846835010183
UDI-Public(01)10846835010183
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P030031
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberD134805
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/05/2026
Initial Report FDA Received Date08/31/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
Patient Age77 YR
Patient SexMale
-
-