• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Inogen Inc. Inogen Rove 4 Portable Oxygen Concentrator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Inogen Inc. Inogen Rove 4 Portable Oxygen Concentrator Back to Search Results
Model Number IO-401
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Codes Dyspnea (1816); Low Oxygen Saturation (2477)
Date of Event 08/04/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient's battery was not lasting as long with a full charge.As a result, the unit lost power and completely shut down causing the patient to have shortness of breath due to no oxygen from the unit.Troubleshooting did not resolve the issue, but it was determined that the battery needed to be replaced.A replacement battery was sent to the patient and it is working as intended for them.Because the battery is an accessory to the device, the unit is listed below, but the issue remains with the battery.
 
Additional Manufacturer Narrative
B3 date of event is estimated.The battery is considered as an accessory and was not returned for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Inogen Rove 4 Portable Oxygen Concentrator
Common Device Name
Inogen Rove 4 Portable Oxygen Concentrator
Manufacturer (Section D)
Inogen Inc.
600 shiloh rd
plano TX 75074
Manufacturer (Section G)
Inogen Inc.
600 shiloh rd
plano TX 75074
Manufacturer Contact
greg porter
600 shiloh rd
plano, TX 75074
MDR Report Key50531547
Report Number3004672275-2026-00276
Device Sequence Number9910724
Product Code CAW
UDI-Device Identifier00817131020414
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K222086
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 08/31/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberIO-401
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/07/2026
Initial Report FDA Received Date08/31/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/02/2026
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient SexUnknown
-
-