| Model Number |
CF-H190L |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the colonovideoscope for a separate issue.During the initial evaluation of the returned device, scope error (e217) was observed.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is unknown.Additional information will be provided once available.Based on the final investigation of the device the initial evaluation findings were confirmed.A root cause could not be identified.Based on the results of the investigation, the most probable cause of the malfunction scope error (e217) was due to data error of scope identification (id).Olympus will continue to monitor the field performance of this device.Should any additional relevant information become available, a supplemental report will be submitted.
|
| |
|