Zimmer Biomet, Inc. Alliance®; Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
|
Back to Search Results |
|
| Catalog Number |
SBGL2000 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the screw disengaged from the blue vault guide handle while the instrument was being used as a glenoid targeting guide, and a second handle was obtained to continue the procedure.During cleaning after the procedure, the screw, which had been safely retained from the patient and operative site, was subsequently lost in the washer.There were no consequences or impact to the patient.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4)
the complaint can be confirmed through the returned product analysis.Fracture analysis has been previously analyzed for this fracture location where overload was identified as the mode of failure.Visual examination of returned product identified signs of use as well as wear and tear.The handle of the device has some knicks and wear marks from use.The body of the sizer handle has some scratches as well.The tip of the device is slightly deformed, and the plunger tip has fractured from the shaft.The fractured plunger tip was not returned with the device.The device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|
|
|