| Model Number |
P1170B |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Baxter received a report that care assist ca300 had brakes not holding.The process step during which this occurred was unknown.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter address: (b)(6).The baxter technician found the casters stems broken and needed to be replaced.Per the baxter service manual the bed should be subject to an effective maintenance program.An annual service of the bed is advised in order to maintain its characteristics and performance.Brake casters should be checked for cuts, wear and quality of tread, etc.And replaced when necessary.Check the brakes to see whether the bed moves when the brake pedals are pressed and repair as necessary
a search of the baxter maintenance records showed baxter performed preventative maintenance on this bed in jun 1, 2026.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the casters to resolve the reported event.Based on this information, no further action is required.
|
| |
|
Search Alerts/Recalls
|