Medline Industries, LP MEDLINE; GLOVE,SURG, SENSICARE PI MICRO,LF,PF,7.5 Back to Search Results
Catalog Number MSG9675
Medical Device Problem Codes Material Separation (1562); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/18/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the gloves got torn when putting them on.".
 
Additional Manufacturer Narrative
The reported concern was tearing when donning; the tear occurred at the wrist, and there was a slight delay in patient care due to frequent glove changes.The account provided both a photo and a sample; the photo depicts a glove positioned on the glove bag, with the cuff torn from the glove and additional material missing from the cuff to the palm, and both gloves underwent visual inspection, dimensional evaluation, grip testing, and donning/doffing assessment.During visual inspection, one glove was already torn and was confirmed to match the glove shown in the submitted photo; during donning, the second glove experienced complete separation of the wrist from the palm, and medline can confirm the reported issue.Medline was unable to determine the root cause behind the reported issue; inspection records confirmed they met established quality requirements, and the quality team will continue to monitor for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
GLOVE,SURG, SENSICARE PI MICRO,LF,PF,7.5
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
nicholas kourkine
three lakes drive
northfield, IL 60093--275
MDR Report Key50533299
Report Number1417592-2026-29085
Device Sequence Number21309922
Product Code KGO
UDI-Device Identifier10888277417304
UDI-Public10888277417304
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K193217
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 08/31/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMSG9675
Device Lot Number311183EB75-3
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/10/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2026
Initial Report FDA Received Date08/31/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-