| Model Number |
1201 |
| Medical Device Problem Codes |
Fracture (1260); Communication or Transmission Problem (2896); Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/21/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient with the neurostimulator presented with a red light displayed on their remote control during ongoing use.Device interrogation identified an initial connection error, which was cleared.Open impedances were observed on electrodes e1 and e2.Multiple impedance checks were performed with the patient in different positions, and the impedance status varied with positioning.Based on the troubleshooting results, a conductor wire fracture was suspected.No troubleshooting resolved the issue.The patient underwent a revision surgery for both the neurostimulator and tined lead on (b)(6) 2026.
|
| |
|
Additional Manufacturer Narrative
|
|
Block d2b:
additional pro code: qon.Block d10:
model number/catalog number: 4101.Serial number: (b)(6).Batch/lot number: ax1t052526.Model/catalog description: 410.Unique identifier (udi) #: (b)(4).Block g4:
additional premarket / 510 (k): p190006.Detailed product investigation has not been provided.We complete a good faith effort to obtain information.Because the product investigation is unknown at this time, we are unable to provide the product investigation and other specific product information.If additional details become available, a supplemental report will be submitted.
|
| |
|