• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION Orca; Endoscope Channel Accessory

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION Orca; Endoscope Channel Accessory Back to Search Results
Model Number SUV-629-50
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that an orcapod 4 was used during procedure performed on an unknown procedure date.During the procedure, the nurse reported that the air/water button leaked continuously throughout the case; requiring her to place a towel beneath the physician to absorb the excess water.The procedure was completed with this device.There were no patient complications reported as a result of this event and the patient's condition is stable.Attempts to obtain additional information regarding the circumstances surrounding this event have been unsuccessful to date.Should additional relevant details become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
B3: date of event: date of event was approximated to 08/01/2026 as no event date was reported.Block d4, h4: the complainant was unable to report the lot number; therefore the manufacture date and expiration date are unknown.Block h6 (device codes): imdrf device code a050401 captures the reportable event of the of air/water valve fluid/blood leak.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Orca
Common Device Name
Endoscope Channel Accessory
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
model farm road
cork, T12 Y K88
EI   T12 YK88
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, 01752
EI   01752
MDR Report Key50534788
Report Number3005099803-2026-03550
Device Sequence Number6040184
Product Code ODC
UDI-Device Identifier08714729981589
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K182275
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/31/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSUV-629-50
Device Catalogue NumberSUV-629-50
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/04/2026
Initial Report FDA Received Date09/01/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-