| Model Number |
CDDRA500T |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
08/12/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient presented with a pocket infection.The physician explanted the entire system, including the implantable cardioverter-defibrillator (icd), right ventricular (rv) lead and right atrial (ra) lead.A new icd, rv lead and ra lead were implanted on (b)(6) 2026.The patient remained in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
Further information was requested but not received.A device history record (dhr) review was performed and review of the sterilization records confirmed normal sterilization cycles for the products.The device met specifications prior to leaving abbott manufacturing facilities.The results of the investigation are inconclusive since the device was not returned for analysis.
|
| |
|
Search Alerts/Recalls
|