| Model Number |
BF-XT190 |
| Medical Device Problem Code |
Image Display Error/Artifact (1304)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the bronchovideoscope when turned on in any direction lines were on the monitor during setup.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for inspection.The root cause could not be identified.Based on the results of the investigation, the customer allegation was confirmed.Additionally, the investigation observed image flickering.It is likely the following led to the malfunction: electrical/electronic problem resulting in component failure.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|