• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT Powered Surgical Solutions MIDAS REX MR8; MOTOR, DRILL, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT Powered Surgical Solutions MIDAS REX MR8; MOTOR, DRILL, PNEUMATIC Back to Search Results
Model Number MR8-AVA15
Medical Device Problem Code Vibration (1674)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/25/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the milling head has too much play.There was no patient involvement.Additional information was received stating that detached bearings were found during evaluation.
 
Additional Manufacturer Narrative
H3: product analysis: evaluation of the device determined that milling head has too much play due to detached bearings.It was also noted tube laser markings are faded, and the color band is faded.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDAS REX MR8
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
MDT Powered Surgical Solutions
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT Powered Surgical Solutions
4620 n beach st
fort worth TX 76137
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
+1(763)526-5677
MDR Report Key50536558
Report Number1625507-2026-07666
Device Sequence Number9865690
Product Code HBB
UDI-Device Identifier00643169912670
UDI-Public(01)00643169912670(11)210423
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K183515
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMR8-AVA15
Device Catalogue NumberMR8-AVA15
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/15/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2026
Initial Report FDA Received Date09/01/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/23/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-