the iab was returned with the membrane completely unfolded and blood on the exterior of the catheter with the one way valve attached.The inner lumen was found occluded with dried blood.The occlusion was unable to be cleared.A sensor output test was performed and the sensor was found to be within specification.The reported event cannot be confirmed by the evaluation.
|