• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Datascope Corp. Sensation Plus 8Fr. 50cc IAB with Accessories, US Only; System, Balloon, Intra-Aortic And Control

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Datascope Corp. Sensation Plus 8Fr. 50cc IAB with Accessories, US Only; System, Balloon, Intra-Aortic And Control Back to Search Results
Model Number 0684-00-0576-01U
Medical Device Problem Codes Obstruction of Flow (2423); Device Sensing Problem (2917)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/07/2026
Type of Reportable Event Malfunction
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Updated fields: a2, a3, a4, b4, b7, d10, g3, g6, g7, h2, h3, h6(problem code, type of investigation, investigation findings, component codes, investigation conclusion), h11 corrected fields: g4,a1 need revert to blank, d8 need revert to blank the iab was returned with the membrane completely unfolded and blood on the exterior of the catheter with the one way valve attached.The inner lumen was found occluded with dried blood.The occlusion was unable to be cleared.A sensor output test was performed and the sensor was found to be within specification.The reported event cannot be confirmed by the evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Sensation Plus 8Fr. 50cc IAB with Accessories, US Only
Common Device Name
System, Balloon, Intra-Aortic And Control
Manufacturer (Section D)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer Contact
ewelina chorlowska
15 law drive
fairfield, NJ 07004
MDR Report Key50538545
Report Number2248146-2026-0001487
Device Sequence Number17630179
Product Code DSP
UDI-Device Identifier10607567109633
UDI-Public(01)10607567109633(10)3000511517(17)281105(11)251105
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K133074
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0684-00-0576-01U
Device Catalogue Number0684-00-0576-01U
Device Lot Number3000511517
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/31/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2026
Supplement Date Received by Manufacturer09/09/2026
Initial Report FDA Received Date09/01/2026
Supplement Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/05/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
IABP
Patient Age56 YR
Patient SexMale
Patient Weight250 LB
-
-