| Catalog Number |
09P69-01 |
| Medical Device Problem Code |
Sparking (2595)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer observed a spark from a wire behind the stain baths of the alinity hs processing module while removing the baths for cleaning that occurred overnight.After replacing the baths, the customer observed the spark again from the same wire.No impact to patient management and no impact to user safety was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The field service representative was attempting to replicate the scenario when troubleshooting the issue but was unable to replicate.Further troubleshooting service found cables trapped around stain bath drawer rails and removed the cables to inspect for damage but found no damage.Service adjusted the cables, but no parts were replaced.The instrument was verified to be functional after cables were adjusted.The instrument service history for the alinity hs processing module, serial number (b)(6) verifies no subsequent complaints have been reported related to observed electrical sparks.A review of complaint and trending data for the alinity hs processing module did not identify any similar issues related to the current issue.Review of the manufacturing documentation did not identify any non-conformances or potential non-conformances associated with the complaint issue.Based on the available information, no systemic issue or deficiency of the alinity hs processing module for serial number (b)(6) was identified.
|
| |
|