Product complaint # (b)(4)
date sent to the fda: 10/2/2026
this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found
additional information: d9, h3, h6
h3 investigational summary: the product was returned to ethicon for evaluation.One open box with twenty representative unopened samples was received for analysis.Product code 8718h.Upon initial inspection of the samples, no external damages were observed on the external packets.As per the sampling plan, a visual inspection was performed on thirteen samples, the packets were opened, and the swage and attachment area were noted to be as expected.The tip and body of the needles were examined under magnification, and no defects or needle breakage were found.In addition, the samples were tested by resistance test to needle and met the requirements.As part of ethicon quality process all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event.To avoid this kind of damage: grasp the needle in an area one-third (1/3) to one-half (1/2) of the distance from the attachment end to the point.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|