Batch review performed on 06 august 2026 stem: quadra-h 01.12.035 quadra stem quadra laterized hap 12/14 s5 (k082792) lot: 130165: (b)(4) items manufactured and released on 13-mar-2013.Expiration date: 28-feb-2018.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Clinical evaluation performed by clinical affairs department a revision surgery was performed 13 years after the primary tha due to femoral stem loosening.The available radiographic image confirms the reported event, showing areas of bone rarefaction around the proximal stem associated with radiolucent lines, findings compatible with loosening.Aseptic loosening is a well-documented complication of total hip arthroplasty and may have a multifactorial or unclear etiology.Given the long interval from primary implantation, multiple biological and mechanical factors may have contributed over time.Based on the limited information available, the precise mechanism leading to stem loosening cannot be determined.No definitive conclusion regarding the underlying failure mechanism can therefore be reached based on the available evidence.Root cause:aseptic loosening is possible literature described adverse event after primary joint arthroplasties and causes are often unknown.There is no indication that any potential issue with the device may have caused or contributed to the event and the investigation does not indicate any potential manufacturing related issue or systemic deficiency.
|